Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00921-1
Product Name/Description Interventional Angiography System

Product codes: INFX8000C; INFX-8000V

ARTG 94739
(Canon Medical Systems ANZ Pty Ltd X-ray system, diagnostic, fluoroscopic, angiographic, stationary, digital)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/10/2024
Responsible Entity Canon Medical Systems ANZ Pty Ltd
Reason/Issue If the SPOT fluoro mode is activated, and the user changes the location of the ROI from the default position then changes the acquisition program, the collimator blades will return to the default position, whilst the display indicates no change.

In the case when the acquisition program is changed the displayed SPOT Fluoro x-ray exposure may be partially covered by the static image. This may lead to unintentional X-ray exposure to the region outside the selected ROI.
Recall Action Product Defect Correction
Recall Action Instructions A software update is being developed to correct the problem. Customers will be contacted when it is available.

Until the software update is available customers are advised to not use the Spot Fluoro function. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.
Contact Information 1300 655 155 - Canon Medical Systems ANZ