Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00863-1
Product Name/Description Oxygen / Nitrous Oxide Circuit

Multiple Part Numbers, Lots Numbers and Expiry Dates

ARTG 210997
(Flexicare Australia Pty Ltd - Breathing circuit, anaesthesia, single use)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/10/2024
Responsible Entity Flexicare Australia Pty Ltd
Reason/Issue Flexicare Medical Ltd has become aware of a potential problem that may arise from the disconnection of a component from the NRV valve which could enter a patient’s airway, if during use the circuit becomes detached and is subsequently reconnected.

A design correction has been implemented between the T-piece and the circuit components; Flexicare Medical Ltd is recalling product manufactured prior to this implementation.
Recall Action Recall
Recall Action Instructions Customers are to quarantine devices from the affected lots. Once quarantined these are to be returned with the customer acknowledgement form. Upon receipt, Flexicare Medical Ltd will issue replacement or credit to customers.
Contact Information 0450 977 007 - Kenny Ho, Flexicare Medical