Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00819-1
Product Name/Description Xpert Xpress CoV-2/Flu/RSV plus. An in vitro diagnostic medical device (IVD)

Part number: XP3COV2/FLU/RSV-10

Batch numbers: 1001420157, 1001421769, 1001423028, 1001423031

Lot numbers: 19108, 19208, 19214, 19219

Expiry dates: 09-Mar-25, 23-Mar-25, 30-Mar-25, 30-Mar-25

ARTG 231513
(Cepheid Holdings Pty Ltd - Multiple-viruses IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/09/2024
Responsible Entity Cepheid Holdings Pty Ltd
Reason/Issue There have been reports that some Cepheid customers have encountered a higher number of Error E5007 "Probe Check Too Low errors" than would typically be expected when performing these tests as described in the product list and their intended use specimen types. An Error E5007 yields a nondeterminate result (Error).
Recall Action Product Defect Correction
Recall Action Instructions If a user experiences an E5007 error when using a cartridge from the product list, they should repeat testing on another cartridge.

If another Error E5007 error occurs, or if the user has been experiencing intermittent E5007 errors on tests from these lots, Cepheid will replace any remaining tests in customers' inventory from affected lots.
Contact Information 1800 107 844 - Ricky Widjaja