Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00756-1
Product Name/Description ADVIA Centaur anti-TPO and Atellica IM Anti-Thyroid Peroxidase. An in vitro diagnostic medical device (IVD).

Used on ADVIA Centaur XP and XPT Systems and Atellica IM Analysers

ADVIA Centaur anti-TPO
100 Test Lot numbers: 26053337 and 26054337
500 Test Lot numbers: 26051337 and 26052337

Atellica IM Anti-Thyroid Peroxidase
100 Test Lot number: 26098336
500 Test Lot number: 26099336

ARTG 175697
(Siemens Healthcare Pty Ltd- Clinical chemistry autoimmune IVDs)
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 28/08/2024
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthineers has confirmed a positive bias in the patient results at and below the cut-off of 60 U/mL with the impacted lots on the ADVIA Centaur XP, ADVIA Centaur XPT and Atellica IM analyzer. Falsely elevated aTPO results may be generated as a result of this problem.
Recall Action Recall
Recall Action Instructions Customers should discontinue use of and discard the impacted lots.
Contact Information Matt Kay on 0438 399 217 or the Siemens Customer Care Center on 1800 310 300.