Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00676-1
Product Name/Description IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent. An in vitro diagnostic medical device (IVD)

IGM reagent cartridge: 1 x 15.2 mL

IGM compartment A: 7.6 mL

IGM compartment B: 7.6 mL

Reference number: 447610

Lot numbers: M307362, M309410

Expiry dates: 31May2025, 31Jul2025

ARTG 213974
(Beckman Coulter Australia Pty Ltd - Clinical chemistry enzyme IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/08/2024
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue IMMAGE Immunochemistry Systems IGM Immunoglobulin M reagent lots M307362 and M309410 are not meeting the sensitivity statement in the IFU:
“Sensitivity is defined as the lowest measurable concentration which can be distinguished from zero with 95% confidence. Sensitivity for IGM determination is 4.2 mg/dL.”
Recall Action Recall
Recall Action Instructions Customers should:
- Discontinue use of affected IMMAGE Immunochemistry Systems IGM Immunoglobulin M reagent lots
- Review the Safety Data Sheet (SDS) and discard according to their laboratory’s Standard Operating Procedures and/or local regulations. The SDS is available on www.beckmancoulter.com
- Contact Beckman Coulter Customer Support Centre or their local Beckman Coulter representative to organise replacement stock
Contact Information 1800 060 881 - Customer Support Centre