Type of Product |
Medical Device |
TGA Recall Reference |
RC-2024-RN-00614-1 |
Product Name/Description |
mo-vis Scoot Control
Product Identifiers: M-P015-65; M-P015-66
Serial Numbers: 3114, 3148-3154 and 3170-4385
ARTG 383810 (Permobil Australia - Wheelchair accessory, steering/control unit) |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
24/07/2024 |
Responsible Entity |
|
Reason/Issue |
In rare cases, the lever of the Scoot Control throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself.
|
Recall Action |
Recall |
Recall Action Instructions |
Customers are advised that until replacement of the affected unit:
Do not over-torque the bolt when adjusting the placement of the throttles. Only 1 Nm should be applied. - There is a likelihood that a mechanical malfunction will occur. Be cautious during use and you must be aware that the on/off switch must be used immediately in case of the throttle being stuck. - If the device malfunction occurs, the device should not be used anymore and should be sent back to the distributor.
The distributor will be in contact to arrange for a device exchange. |
Contact Information |
1300 845 483 - Permobil Australia |