Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00600-1
Product Name/Description Proteus XR/a Table Dolly Usage and Product Deinstallation Instructions

GTIN: 00840682120777

All Systems

ARTG 1676117
(GE Healthcare Australia Pty Ltd - X-ray system, diagnostic, general-purpose, stationary, digital)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/07/2024
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE HealthCare has become aware of an event involving a 3rd party service organisation that occurred while de-installing a Proteus XR/a table. The table dolly was not utilised properly, resulting in a minor injury to the person using the service tool.

During deinstallation of a Proteus patient table, the transport dolly bolts securing a dolly to the table assembly stripped when a third-party field service engineer (FSE) lowered the assembly to the floor. The intended deinstall process is to first remove the tabletop from the table assembly, and the dolly used was originally an install dolly which was not anticipated to be saved and eventually used for deinstallation.

This problem is regarding a service tool utilised to install or deinstall the x-ray table portion of the equipment and has no effect on clinical use of the equipment.
Recall Action Product Defect Correction
Recall Action Instructions Customers can continue to use the Proteus XR/a as intended for clinical use.

De-installation Instructions have been released for the product. If customers have retained one of the service dollies or intend to contract a 3rd party to deinstall their Proteus XR/a, they must utilise the De-installation Instructions.

Customers are to pay attention to Section 1.4.1 of the instructions regarding the length of bolts specified for use with the dolly and ensure these specified bolts are used as required.

Customers can find the de-installation instructions on the product documentation portal by searching for Proteus XR/a De-Installation Manual or instruction number 5273026-8EN. https://www.gehealthcare.com/support/manuals.

This action has been closed out on 23/08/2024
Contact Information 1800 659 465 - GE HealthCare Service