Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00597-1
Product Name/Description Philips SENSE XL Torso (1.5T and 3.0T) Coils

SENSE XL TORSO COPIL 1.5T
Model: 453567141882, 453567141883

SENSE XL TORSO COIL 1.5T Mk2
Model: 453567502281

SENSE XL TORSO COIL 3.0T
Model: 456567394942, 456567394943, 456567394945

ARTG 425091
(Philips Electronics Australia Ltd - Coil, MRI, radio frequency)
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 16/07/2024
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips has identified a problem in the SENSE XL Torso (1.5T and 3.0T) coils which may result in localised heating during the scan, resulting in potential for a heating sensation and/or 1st, 2nd or 3rd degree burns around the area covered by the coil.

Philips has received 64 complaints of coil heating events, including 52 reports of patient harm associated with this issue as of April 2024.
Recall Action Product Defect Correction
Recall Action Instructions Customers are able to continue use of the coils in accordance with the Instructions for Use (IFU).
1. Avoid First Level Operating Mode/High SAR scans
2. Use dedicated pads - Always use dedicated pads and mattresses provide with the coils.

In addition:
1. Avoid positioning the coil close to the bore
When using the SENSE XL (1.5T and 3.0T) Torso Coils, ensure the anterior portion of the coil is positioned greater than 2 inches (5 cm) away from the bore.

2. Do not exceed 45 minutes of examination time (excluding setup time) for a single patient.

Philips is also initiating design of an improved coil. An update on the development of this plan is expected to be relayed by the end of 2024.
Contact Information 1800 251 400 - Philips Service Delivery Team