Type of Product |
Medical Device |
TGA Recall Reference |
RC-2024-RN-00552-1 |
Product Name/Description |
Medtronic CareLink 2090 Programmer Autonomous Cursors
PRGM 2090 PROGRAMMER-PORTABLE CRM Model number: 2090 Serial Number: Devices with serial number prefixes PKK0 and PKK1
PROGRAMMER 2090W TELEMETRY C Model number: 2090W Serial number: Devices with serial number prefixes PKK0 and PKK1
ARTG 116038 (Medtronic Australasia Pty Ltd - Medtronic carelink programmer, model 2090 - Pacemaker programmer) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
10/07/2024 |
Responsible Entity |
|
Reason/Issue |
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Through 11 June 2024, Medtronic has received 23 reports of autonomous cursor behaviour, with 2 reports of unintended therapy delivered when the programmer was not under the control of trained personnel during a patient session. There have been no reports of permanent harm or death associated with this behaviour. |
Recall Action |
Product Defect Correction |
Recall Action Instructions |
Medtronic will contact affected customers to assist with the software updates and assessing the proper function of the affected devices.
If needed, Medtronic representatives will assist with returning programmers for repair or replacement. |