Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00530-1
Product Name/Description CADD Medication Cassette Reservoirs

Product Descriptions: RESERVOIR, CASSETTE, 100ML 12/BX, RESERVOIR, CASSETTE, 100ML, FS, YELLOW 12/BX, RESERVOIR, CASSETTE, 50ML, FS 12/BX, RESERVOIR, CASSETTE, 100ML, FS 12/BX

Model Numbers: 21-7002-24, 21-7300-24, 21-7301-24, 21-7302-24

Multiple lot numbers

ARTG 400443
(Smiths Medical Australasia Pty Ltd - Infusion pump cassette)
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 3/07/2024
Responsible Entity Smiths Medical Australasia Pty Ltd
Reason/Issue Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could cause a medication leakage that could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.

To date no serious injuries or deaths have been reported to Smiths Medical for this problem.
Recall Action Recall
Recall Action Instructions Customers are to:
Review inventory of CADD Medication Cassette Reservoirs to confirm if they have any of the affected models/lots. Dispose of these according to their facility's policy.

A credit will be issued once Smiths Medical receives the acknowledgement form.
Contact Information 1800 654 949 - Smiths Medical Customer Service