Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00466-1
Product Name/Description CADD Infusion System Infusion Sets

List Numbers: 21-7300-24, 21-7309-24, 21-7322-24, 21-7323-24, 21-7359-24, 21-7391-24, 21-7394-24

Multiple product descriptions and Batch numbers

ARTG's: 397091 & 400443
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/06/2024
Responsible Entity Smiths Medical Australasia Pty Ltd
Reason/Issue Smiths Medical has identified additional lots of CADD Infusion System Infusion Sets that are potentially affected as detailed in the previous action of RC-2023-RN-00529-1 for problem 1.

Problem 1: Manufacturing variations may cause the green CADD Flow Stop arm to compress and partially occlude the tubing before clinical use. This may result in under delivery, despite the pump displaying that the infusion is running properly.
Recall Action Recall
Recall Action Instructions Smiths Medical are advising pharmacists, clinicians and patients to use alternate CADD infusion sets for life sustaining medications.

Smiths Medical will replace impacted stock. Impacted stock should be returned for credit.

This action has been finalised on 20/8/2024.
Contact Information 1800 654 949 - Customer Service