Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00446-1
Product Name/Description UNIMAX Detachable Endo Pocket

Catalogue Numbers: SB936, SB957 and SB979

Lots for SB936: 6252304017, 6252304033, 6252305087

Lots for SB957: 6252207180, 6252303004, 6252303018, 6252304025, 6252304034

Lots for SB979: 6252207182, 6252303007, 6252304026, 6252304035, 6252304047

ARTG 202436
(ConMed Linvatec Australia Pty Ltd - Bag, tissue extraction)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/06/2024
Responsible Entity ConMed Linvatec Australia Pty Ltd
Reason/Issue ConMed Australia have received reports that the black heat-shrink tubing material supporting the ring at the top of the bag may detach during use.
There is a risk it may fall into the surgical site and retrieval could cause a delay in the surgical procedure.

ConMed Australia has received no reports of any patient injury or death associated with this problem.
Recall Action Recall
Recall Action Instructions -Customers are to check for affected Lots on their inventory, discontinue use of the affected devices immediately and return them to ConMed Australia for replacement.

-Complete and return the attached Reply Form to ConMed Australia by fax to 1800 238 298 or email to RegulatoryAU@conmed.com and ConMed will contact affected customers to organise the return and replacement of the products.

This action has been closed out on 14/08/2024
Contact Information 1800 238 238 - RegulatoryAU@conmed.com - ConMed Australia Customer