Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00418-1
Product Name/Description DxI 9000 Access Immunoassay Analyser. An in vitro diagnostic medical device (IVD)

Software: 1.17.0 and below

REF: C11137

Multiple assays and reagent part numbers

ARTG 177999
(Beckman Coulter Australia Pty Ltd - Instrument/analyser IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/05/2024
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue If a sample test result relative light units (RLUs) are between the RLU value of the calibration curve math model parameter A and the S0 calibrator level mean RLUs, then the DxI 9000 Access Immunoassay Analyser will report No Result rather than 0.

There may be a delay in reporting test results which could lead to a delay in patient treatment.

This issue affects approximately 0.04% of Access assay patient sample test results. The Access Sensitive Estradiol assay has the highest rate of occurrence (approximately 2.2%).
Recall Action Product Defect Correction
Recall Action Instructions If No Result is observed for a patient test, the operator should:
• Rerun the sample test on either the same or a different analyser.
• Recalibrate the assay.
• Repeat the sample test on a different immunoassay platform if one is available.

Beckman Coulter has identified the root cause of this problem and will implement the correction with a future software release.
A Beckman Coulter service representative will contact customers to schedule the system software upgrade as soon as it becomes available.
Contact Information 1800 060 881 - Customer Support Centre