Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00361-1
Product Name/Description Zenition 70 systems

Model Number: 718133

ARTG 235144, 405770
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 8/05/2024
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue There is a possibility of unintended radiation exposure while using the wired footswitch or handswitch.

The requirement from the standard states that X-ray stop time should not exceed 100 milliseconds once the footswitch or handswitch is released to stop the X-ray. However, there might be a possibility when a Zenition 70 system is used in 2fps subtract mode for a vascular surgery procedure, the time to stop X-ray might exceed by 18 milliseconds over and above the 100 milliseconds.
Recall Action Product Defect Correction
Recall Action Instructions A Philips representative will contact all affected customers to schedule a Field Service Engineer visit to update the software free of charge.
Contact Information 1800 251 400 - Philips Service Delivery Team