Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00304-1
Product Name/Description Philips Spectral CT7500

Product code: 728333

Device Identifier numbers: 00884838101111, 00884838111103

ARTG 423851
(Philips Electronics Australia Ltd - X-ray system, diagnostic, computed tomography, full-body)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/04/2024
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue The HIB-CIB cable on the NOAH couch may wear unexpectedly over time.

During vertical couch motion, if the HIB-CIB cable has been damaged, the system may engage the Emergency Stop intermittently.

If the Emergency Stop is engaged both motion and X-rays will stop.

If this occurs during a scan it could result in a partial set of images, and a patient rescan may be required.

Philips has not received any reports of injury or serious harm associated with this problem to date.
Recall Action Product Defect Correction
Recall Action Instructions Philips will contact the affected customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the impacted site and install the solution to resolve the problem.

In the meantime, customers may continue to use the Philips CT system(s) in accordance with the intended use.

If they experience the Emergency Stop opening intermittently during vertical couch motion, they must contact the local Philips representative.
Contact Information 1800 251 400 - Philips Service Delivery Team