Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00175-1
Product Name/Description DxI 9000 Access Immunoassay Analyser. An in vitro diagnostic medical device (IVD)

Multiple serial numbers

Product Code/Catalogue Number: C11137

Software: 1.16.2 and below

ARTG 177999
(Beckman Coulter Australia Pty Ltd - Instrument/analyser IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/03/2024
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue In the DxI 9000 Access Analyzer Instructions for Use (IFU section “System Configuration”) the operator may configure a sample and select a reporting unit. If the user fails to follow the Instructions for Use (IFU) and does not select a unit of measurement (e.g., the field is left blank), the analyser will report the results as 0 mmol/L.

For example, if plasma is selected for the Access TSH assay and the unit field is left blank, the analyser will report the TSH results as 0 mmol/L.
Recall Action Product Defect Correction
Recall Action Instructions Customers should:
• Always configure tests as instructed in the DxI 9000 Access Analyser Instructions for Use (IFU).
• Confirm and review the appropriate units of measure were selected if new sample types have been configured.
• If a unit of measure was left blank:
o Select the appropriate unit of measure for the sample type.
o Perform a retrospective review of applicable test results.
• Repeat any tests as necessary.
Beckman Coulter has identified the root cause of this issue and will implement a correction with a future software release. A Beckman Coulter service representative will contact customers to schedule the software upgrade when it is available.
Contact Information 1800 060 881 - Customer Support Centre