Type of Product |
Medical Device |
TGA Recall Reference |
RC-2024-RN-00161-1 |
Product Name/Description |
Chiba Biopsy Needle
Reference part number (RPN): DCHN-22-20.0-U
Order number (GPN): G04338
Lot number: NS15808547
ARTG 138559 (William A Cook Australia Pty Ltd - Biopsy Access Needle - Needle, aspirating, single use) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
28/02/2024 |
Responsible Entity |
|
Reason/Issue |
Cook Medical is recalling a specific lot (NS15808547) of Chiba Biopsy Needles. The affected lot may have packaging seals that do not meet peel strength specifications.
This creates a potential of a sterile barrier failure that could result in compromised device sterility.
Potential adverse events include increased procedural time and infection, potentially being life-threatening and/or requiring medical/surgical intervention.
Cook Medical has not received any customer complaints related to the adverse patient effects listed above for the affected lot.
However, compromised device sterility may go undetected by the user. |
Recall Action |
Recall |
Recall Action Instructions |
Customers must examine their inventory immediately for affected products, quarantine and cease use and return them to Cook Medical. |
Contact Information |
1800 777 222 - Cook Medical Customer Relations |