Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00143-1
Product Name/Description Low RNA Reagent Cassette. An in vitro diagnostic medical device (IVD)

Catalogue number: 40331 version: 03

Batch numbers: C298133RL, C299183RL, C297907RL, C298045RL

Date of expiry: 26-Sep-2024, 15-Oct-2024, 20-Sep-2024, 25-Sep-2024

ARTG: 236377
(AusDiagnostics Pty Ltd - Multiple infectious organism IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/02/2024
Responsible Entity AusDiagnostics Pty Ltd
Reason/Issue AusDiagnostics has identified a potential raw material problem leading to background noise / non-specific reactivity of the Parainfluenza virus (PIV) type 1, 2 or 3 targets, which the software is interpreting as positive / detected.

To date, there have been no reports of adverse events related to this issue.
Recall Action Recall
Recall Action Instructions Customers are to quarantine and dispose of the remaining affected batches.
AusDiagnostics Pty Ltd will provide replacement stock to customers once the customer acknowledgement form is received.
Contact Information 02 9698 8030 - AusDiagnostics