Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00111-1
Product Name/Description Xray Screening table handle

S/N: MQC3548CA001

ARTG 141008
(Shimadzu Medical Systems Oceania Pty Ltd - X-ray system, diagnostic, fluoroscopic, general-purpose, stationary, analogue)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/02/2024
Responsible Entity Shimadzu Medical Systems Oceania Pty Ltd
Reason/Issue It has been found that the Exposure Button on the handle of a control panel manufactured during a specific period may not be fully released from the ON state after releasing the button.
Recall Action Product Defect Correction
Recall Action Instructions Shimadzu will replace the handle with an improved handle so that the Exposure Button is
released from the ON state without fail. Our Shimadzu service department will contact customers shortly to schedule this work.

Until such time that the work can take place, the following is recommended:

1. Perform daily inspections according to the Flexavision Operation Manual.

2. If any failures are observed, stop using the equipment and contact Shimadzu Medical Systems Service department.

This action has been closed out on 16/4/2024.
Contact Information 1800 819422 - Shimadzu service number