Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2024-RN-00059-1
Product Name/Description The Alinity s System. An in vitro diagnostic medical device (IVD)

List Number: 06P16.01

Multiple serial numbers

ARTG: 285249
(Abbott Australasia Pty Ltd Diagnostic Division - Instrument/analyser IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/01/2024
Responsible Entity Abbott Australasia Pty Ltd Diagnostic Division
Reason/Issue Abbott has identified an instance of a hardware issue (out-of-specification standoffs) associated with the assembly of the Alinity System incubation track.
This may lead to Reaction Vessel (RV) jams and/or splashing within the RV as it transitions from the incubation track to the process path during instrument operation. Splashing within an RV has the potential to affect test results.
No Adverse events due to this issue have been reported in Australia.
Recall Action Product Defect Correction
Recall Action Instructions Customers are requested to continue operating and maintaining the Alinity System as described within the System Operations Manual. There are no additional necessary actions to be taken by customers.

Abbott representatives will inspect all Alinity s Systems to identify any standoffs out of specification. If any standoffs are identified as incorrect, Abbott will work with impacted customer sites to schedule the required service activities to replace these standoffs.
Contact Information 1800 816 696 - Abbott Customer Support