Type of Product Medical Device
TGA Recall Reference RC-2023-RN-00734-1
Product Name/Description The Gallant, Entrant, and Neutrino NxT HV devices

Model numbers: CDVRA500Q, CDDRA500Q, CDHFA500Q, CDVRA600Q, CDDRA600Q, CDHFA600Q, CDVRA300Q, CDDRA300Q, and CDHFA300Q

ARTGs: 336475, 341252, 336476, 341253, 341254, 336479, 336477, 336478, 336480
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/08/2023
Responsible Entity Abbott Medical Australia Pty Ltd
Reason/Issue Abbott is informing customers of a rare potential for a Bluetooth (BLE) circuit component issue on a subset of Gallant, Neutrino, and Entrant Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronisation Therapy Defibrillators (CRT-Ds) manufactured prior to April 2022.

This issue has been associated with a risk of loss of Bluetooth communication (and therefore a loss of remote monitoring), higher than normal current consumption, and reduced device longevity.

If a device experiences this issue, primary device functions, including pacing, sensing, shock delivery, and inductive telemetry, remain available during the period of remaining battery life.
Recall Action Product Defect Correction
Recall Action Instructions Abbott advises customers:
· Prophylactic device replacement is NOT recommended as the new firmware version pr00.10.87.04 eliminates the potential for loss of therapy between follow-ups due to unrecognized decreased device longevity
· Determine the firmware version of devices followed at your clinic by reviewing the instructions in the customer letter
· For patients with firmware version pr00.10.87.00 or with firmware version undetermined, upgrade devices to device firmware version pr00.10.87.04 by interrogating patients in-clinic with Merlin PCS 3650 programmer Model 3330 software version 25.4.1 rev 1 or higher or any Merlin 2 PCS MER3700 programmer
· Prioritise in-clinic firmware upgrade for the specific devices from the 1,500 device sub-group listed in the customer letter
· For remaining patients, follow-up in-clinic to complete the firmware upgrade
· If a device experiences a loss of Bluetooth communication, contact Abbott Technical Support for troubleshooting.

This action has been closed out on 20/11/2023.