Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2023-RN-00726-1
Product Name/Description Pentacam AXL (Type 70100) and Pentacam AXL Wave (Type 70020)

Serial numbers: 70020 0731 2290, 70020 9731 2290, 70020 7041 2221 and 70020 9631 2290

ARTG 175820
(Designs for Vision Aust Pty Ltd - Corneal topography system)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/08/2023
Responsible Entity Designs for Vision Aust Pty Ltd
Reason/Issue One of the suppliers did not meet the specifications for an optical component inside the devices. An insufficient anti-reflective coating was found, which under certain circumstances can lead to an additional axial length signal. If the correct axial length signal is weaker (e.g. due to an off-centre measurement or a cataract), an incorrect axial length value can be displayed. OCULUS identified the problem within the production process.

There have been no reports of physical injury or serious negative consequences.
Recall Action Product Defect Correction
Recall Action Instructions Customers are to Immediately stop using the affected device for axial length measurement. Mark the affected device with a note or take out of service.
OCULUS will replace or repair the affected devices. A representative will contact customers to coordinate the further procedure (such as exchange, repair, supply of loan devices).

This action has been closed out on 07/12/2023
Contact Information 1800 225 307 - Designs for Vision