Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2023-RN-00678-1
Product Name/Description Philips Spectral CT7500

Product code: 728333

Device identifier: 00884838101111

Serial numbers: 396001, 396003 and from serial number 10014 through serial number 10127

ARTG 321687
(Philips Electronics Australia Ltd - X-ray system, diagnostic, computed tomography, full-body)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/07/2023
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue For any scan type, if the Auto Voice and Breathing Lights buttons on the All Parameters tab are selected, it is possible that the automatic breathing instructions given by the system will be mismatched. The audio instructions to “Breath-In” are given by the system, but the “Breath-Out” Light Emitting Diode (LED) will light up. The “Breath-Out” audio instructions are also reversed with the “Breath-In” LED. The mismatch between the audio and visual breathing instructions may lead to patient confusion potentially resulting in a patient rescan being required.
Recall Action Product Defect Correction
Recall Action Instructions Philips will contact customers to schedule a time for a Field Service Engineer to install the solution to resolve the issue.
In the meantime, customers are to deselect the Auto Voice and Breathing Lights buttons in All Parameters tab for the selected scan and provide the correct breathing instructions manually. The Plan Window and Parameters chapter in section 4 of the IFU has additional information.

This action has been closed out on 15/02/2024
Contact Information 1800 251 400 - Philips Service Delivery Team