Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2023-RN-00540-1
Product Name/Description Philips Big Bore RT, Brilliance CT Big Bore and Brilliance CT Big Bore Oncology (software version 4.8.0)

Product Code: 728242, 728244, 728243

UDI: (01)00884838095168, (01)00884838059450, (01)00884838059450

ARTG 321687
(Philips Electronics Australia Ltd - X-ray system, diagnostic, computed tomography, full-body)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 19/06/2023
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Multiple software issues affecting Philips Big Bore RT and Upgrade Brilliance CT Big Bore systems have been identified. These issues may cause the system to shut down unexpectedly or to generate unsatisfactory images, which may result in the need to perform a rescan.
Recall Action Product Defect Correction
Recall Action Instructions Philips will contact customers to schedule a site visit to install a software update to resolve the issues.

In the interim, customers may continue to use the system(s) in accordance with the intended use. Customers are to refer to the customer letter (supplied by the sponsor to affected customers) for specific details regarding the issue descriptions and recommendations on actions to be taken.

This action has been closed out on 26/07/2024
Contact Information 1800 251 400 - Philips Service Delivery Team