Type of Product Medical Device
TGA Recall Reference RC-2022-RN-01530-1
Product Name/Description OAPL FACE MASK DISPOSABLE 3PLY EARLOOP

Batch number: 205029

Date of manufacture: 05/2020

ARTG 333823 - Cancelled as of 11/08/2022
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 14/12/2022
Responsible Entity Orthotic & Prosthetic Centre Pty Ltd
Reason/Issue The Therapeutic Goods Administration (TGA) has undertaken a post-market review of face masks included in the Australian Register of Therapeutic Goods (ARTG).

Orthotic & Prosthetic Centre (OPC) Pty Ltd has supplied batches of OAPL Face Mask Disposable 3 Ply Earloop which when tested by TGA Laboratories, failed fluid resistance testing at level 2 (120mmHg).

The TGA has cancelled the device from the ARTG under the request of the Sponsor in accordance with section 41GL(d) of the Therapeutic Goods Act 1989 (the Act).
Recall Action Product Defect Alert
Recall Action Instructions The TGA does not require the face masks to be physically recalled (or otherwise removed) from the market.
OAPL Face Mask Disposable should not be used in that sort of clinical or patient/caregiving scenario including hospital wards, operating theatres, general practice, nursing homes or during in-home care / support services where a level 2/3 fluid resistance is required

At the user’s discretion or where the use of face masks is mandated under the provisions of a state or territory public health order, OAPL Face Mask Disposable may also be used in non-clinical environments such as domestic households, retail shops, food and beverage outlets, public transport, outdoor settings, etc.
Customers should consider:
• quarantining the product;
• using them in another setting; and/or
• if they do not have alternatives, ensuring that a face shield is used in high risk settings.
This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.
Contact Information 03 9681 9666 - Product Manager