TGA Recall Reference RC-2022-RN-01330-1
Product Name/Description Syngo.via with VB20, VB30, VB40, VB60 and VB70 software

ARTG 371150
(Siemens Healthcare Pty Ltd - Radiology DICOM image processing application software)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/11/2022
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens has become aware that under certain circumstances, distance measurement values in MRI images might be wrong in syngo.via.

In case:
1. Enhanced MRI images with rectangular FoV (Field of View) are created at the MR scanner and sent to syngo.via, and
2. at the syngo.via system, the user rotates the images to display the images in arbitrary orientation and saves them, and
3. afterwards, the user exports the images to another DICOM node using MR Export Rules (interoperability mode)

then distance measurements on the newly created series might be wrong when being performed at systems that receive these new series.
Recall Action Product Defect Correction
Recall Action Instructions Customers are advised to follow the workaround steps outlined in the customer letter (supplied to impacted facilities). Siemens will solve the issue with a software update which is now available.

This action has been closed out on 29/01/2024
Contact Information 1800 310 300 - Siemens Customer Care Centre