Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2022-RN-01118-1
Product Name/Description BiPAP A40

Serial numbers: V2887246677A0, V288724698F57, V28872472285C, V288724777FF1, V28872479968F, V288724808886, V28872485DF2B, V28872492B24C, V28872495C6F3, V288724981D16, V288725010D13, V2887251114CB and V2887251771FD

ARTG : 200289
(Philips Electronics Australia Ltd - Positive airway pressure unit, bi-level)
Recall Action Level Consumer
Recall Action Classification Class I
Recall Action Commencement Date 16/08/2022
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips Respironics has determined that certain devices were built with motor assemblies that could contain non-conforming plastic material.

The patient may be exposed to the following hazards if the non-conforming material is present:
• Exposure to off-gassing not normally present creating a potential biosafety or toxicological hazard.
• Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
Recall Action Recall
Recall Action Instructions Philips will replace all affected devices. Users will receive new A40 Pro units.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.
Contact Information 1800 009 579 - Philips Sleep & Respiratory Care