Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00929-1
Product Name/Description Platinium Implantable Cardiac Defibrillator (ICD) and Resynchronisation Therapy Defibrillator (CRT-D)

PLATINIUM VR 1210
ARTG No: 282817

PLATINIUM VR 1240
ARTG No: 282816

PLATINIUM DR 1510
ARTG No: 282818

PLATINIUM DR 1540
ARTG No: 282819

PLATINIUM CRT-D 1711
ARTG No: 282820

PLATINIUM CRT-D 1741
ARTG No: 282821

PLATINIUM SONR CRT-D 1811
ARTG No: 282822

PLATINIUM SONR CRT-D 1841
ARTG No: 282823
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 9/07/2018
Responsible Entity LivaNova Australia Pty Ltd
Reason/Issue On a subset of Platinium ICD and CRT-D devices, a specific hardware configuration was identified as potentially defective over time, leading to overconsumption, immediately followed by a loss of pacing and sensing capabilities in all cavities. As a result of this loss of sensing capability, the device cannot identify an arrhythmia that would normally require defibrillation shock therapy.

No permanent injury or death has been reported as a result of this issue.
Recall Action Recall
Recall Action Instructions Regarding the medical device recall:
Microport CRM is initiating a removal of the non-implanted Platinium devices that may present with this hardware failure and providing recommendations for managing implanted patients; and

Regarding the implant hazard alert:
In order to mitigate the potential risks associated with the loss of the device pacing and sensing capabilities, Microport CRM recommends physicians follow-up the patients at the periodicity already stated in the implant manual.

This action has been closed out on the 21/11/2018.
Contact Information 03 9799 7444 - LivaNova