TGA Recall Reference RC-2018-RN-00471-1
Product Name/Description ARTIS pheno systems

Catalogue Number: 10849000

ARTG Number: 273952
(Siemens Healthcare Pty Ltd - X-ray system, diagnostic, fluoroscopic, angiographic, stationary, digital)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 18/05/2018
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens has identified an issue with ARTIS pheno systems. If the C-arm is moved manually longitudinal to a fully extended position it may occur rarely, that the system's safety stop mechanism may be activated. As a result, the C-arm of ARTIS pheno stops the movement and can drop up to 20 mm (0.8"). This can happen during an ongoing procedure.
If this issue occurs, the system performs a safety stop. All C-arm movements are blocked and can only be reactivated by a field service engineer.
Recall Action Product Defect Correction
Recall Action Instructions Siemens are currently working on a solution to eliminate the root cause of the problem.
In the interim, users are advised to avoid driving the C-arm into a fully extended longitudinal position. A fully extended longitudinal position is reached when the stand arms are almost completely stretched out.

This action has been closed out on 27/02/2019
Contact Information 1800 310 300 - Siemens Customer Care Centre