Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00297-1
Product Name/Description Radiometer TCM5FLEX/BASIC

All Serial Numbers

ARTG Number:140333
(Radiometer Pacific - Transcutaneous blood gas monitor)
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 13/04/2018
Responsible Entity Radiometer Pacific Pty Ltd
Reason/Issue Radiometer has recently become aware that the Wi-Fi hardware incorporated in the TCM5 monitor is not compliant with respect to the requirements for radiated emission (Radio Equipment Directive 2014/53/EU).
Australian customers do not currently utilise the Wi-Fi option on these devices.
Recall Action Product Defect Correction
Recall Action Instructions Radiometer is advising that they will be contacting affected users to arrange for the software to be upgraded on TCM5 to V1.2.0 which will disable the Wi-Fi hardware.
Customers have been supplied an updated IFU which removes reference to Wi-Fi as part of the correction undertaken for RC-2018-RN-00299-1

This action has been closed out on 25/10/2018