Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00297-1 |
Product Name/Description |
Radiometer TCM5FLEX/BASIC
All Serial Numbers
ARTG Number:140333 (Radiometer Pacific - Transcutaneous blood gas monitor) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
13/04/2018 |
Responsible Entity |
|
Reason/Issue |
Radiometer has recently become aware that the Wi-Fi hardware incorporated in the TCM5 monitor is not compliant with respect to the requirements for radiated emission (Radio Equipment Directive 2014/53/EU). Australian customers do not currently utilise the Wi-Fi option on these devices. |
Recall Action |
Product Defect Correction |
Recall Action Instructions |
Radiometer is advising that they will be contacting affected users to arrange for the software to be upgraded on TCM5 to V1.2.0 which will disable the Wi-Fi hardware. Customers have been supplied an updated IFU which removes reference to Wi-Fi as part of the correction undertaken for RC-2018-RN-00299-1
This action has been closed out on 25/10/2018 |