Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00291-1
Product Name/Description OXOID AST DISC CT1412B – Cefpirome CPO30ug 5x 50 discs. An in vitro diagnostic medical device.

Lot Numbers:1611705, 1693625

ARTG Number: 208273
Thermo Fisher Scientific Australia & New Zealand - Antimicrobial susceptibility testing IVDs
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 3/04/2018
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue An internal technical investigation by Oxoid Limited, part of Thermo Fisher Scientific, has confirmed that the performance of antibiotic in ThermoScientific Oxoid CT1412B CPO30 Cefpirome may decrease when stored over 2°C after 24 months. The product has a registered shelf life of 36 months.
Performance remains unaffected throughout shelf life when stored at -20°C to +2°C. Continued use of lots older than 24 months if stored at 2-8°C, could result in false resistance reporting for Cefpirome.
Recall Action Recall
Recall Action Instructions Thermo Fisher is advising customers to inspect stock for affected product and remove them from further use. Users should destroy/discard all current and unused stock of the affected batches.

This action has been closed out on the 16/10/2018.