Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00277-1 |
Product Name/Description |
Dimension Vista CTNI Sample Diluent and Multi 2 Sample Diluent. An in vitro diagnostic medical device (IVD)
CTNI Sample Diluent Catalogue Number: KD692 Siemens Material Number: 10445202 Lot Number: 7BDA34 Expiry Date: 01-06-2018
Multi 2 Sample Diluent Catalogue Number: KD694 Siemens Material Number: 10483586 Lot Number: 7CDA79, 7DDA70 Expiry Date: 01-09-2018, 01-11-2018
ARTG Numbers: 182774 and 182222 (Siemens Healthcare - Troponin IVDs) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
27/03/2018 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has confirmed that some Dimension Vista Diluent lots may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the Dimension Vista System, due to a reduced volume of diluent being pipetted from the vial. There have been no injuries reported as a result of this issue. |
Recall Action |
Recall |
Recall Action Instructions |
Siemens is advising users to inspect stock and discard any remaining stock of the affected lots. Siemens is not recommending a review of previously generated results.
This action has been closed out on 25/02/2019 |
Contact Information |
1800 310 300 - Siemens Customer Service Support |