Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00277-1
Product Name/Description Dimension Vista CTNI Sample Diluent and Multi 2 Sample Diluent. An in vitro diagnostic medical device (IVD)

CTNI Sample Diluent
Catalogue Number: KD692
Siemens Material Number: 10445202
Lot Number: 7BDA34
Expiry Date: 01-06-2018

Multi 2 Sample Diluent
Catalogue Number: KD694
Siemens Material Number: 10483586
Lot Number: 7CDA79, 7DDA70
Expiry Date: 01-09-2018, 01-11-2018

ARTG Numbers: 182774 and 182222
(Siemens Healthcare - Troponin IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/03/2018
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare Diagnostics has confirmed that some Dimension Vista Diluent lots may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the Dimension Vista System, due to a reduced volume of diluent being pipetted from the vial.
There have been no injuries reported as a result of this issue.
Recall Action Recall
Recall Action Instructions Siemens is advising users to inspect stock and discard any remaining stock of the affected lots.
Siemens is not recommending a review of previously generated results.

This action has been closed out on 25/02/2019
Contact Information 1800 310 300 - Siemens Customer Service Support