Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00269-1
Product Name/Description Medicina Nasogastric Tubes

Product Codes: LG & NGP

Tube sizes: 4Fr, 5Fr and 6Fr

Multiple Lot Numbers

ARTG Number: 276654
(Australasian Medical & Scientific - Tube, feeding, nasoenteral)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/04/2018
Responsible Entity Australasian Medical & Scientific Ltd
Reason/Issue Australasian Medical & Scientific Ltd (AMSL) have been made aware that the measurement markings on certain nasogastric tubes may be blurred and illegible which may result in a tube being inserted at an incorrect length in a patient.
Recall Action Product Defect Correction
Recall Action Instructions AMSL is advising customers to inspect inventory for identified lot numbers and inspect these lots for any units which display blurry or illegible text. Users should quarantine any affected units for return to AMSL and a replacement to be issued.

This action has been closed out on 15/01/2019
Contact Information 02 9882 3666 - Customer Service Support