Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00254-1 |
Product Name/Description |
PTS Panels CHOL+GLU test strips. An in vitro diagnostic medical device (IVD)
Reference Number: 1765
Lot Number: A705 and A707
Expiry Date: August 2018
Unique Device Identification Code: Lot Numbers: A705: (01)00381931765016(17)180813(10)A705 Lot Numbers: A707: (01)00381931765016(17)180828(10)A707
ARTG Number: 252292 (Point Of Care Diagnostics Australia - Clinical chemistry substrate IVDs) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
27/03/2018 |
Responsible Entity |
|
Reason/Issue |
This recall is being conducted due to the potential for loss of activity of the glucose analyte prior to the full stated shelf life (August 2018). Based on the results of internal testing, there is potential for under-recovery of the glucose analyte, and therefore, potential for lower glucose test results. This issue may result in the delay in diagnosis of diabetes. There have been no adverse events reported in conjunction with this product |
Recall Action |
Recall |
Recall Action Instructions |
Point of Care Diagnostics (PTS) is advising users to inspect stock and quarantine any remaing kits of the affected lots. Affected stock is to be destroyed. PTS will issue replacement product or a credit note.
This action has been closed out on the 26/11/2018. |
Contact Information |
1800 640 075 - Point of Care Diagnostics |