Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00254-1
Product Name/Description PTS Panels CHOL+GLU test strips. An in vitro diagnostic medical device (IVD)

Reference Number: 1765

Lot Number: A705 and A707

Expiry Date: August 2018

Unique Device Identification Code:
Lot Numbers: A705: (01)00381931765016(17)180813(10)A705
Lot Numbers: A707: (01)00381931765016(17)180828(10)A707

ARTG Number: 252292
(Point Of Care Diagnostics Australia - Clinical chemistry substrate IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/03/2018
Responsible Entity Point Of Care Diagnostics Australia Pty Ltd
Reason/Issue This recall is being conducted due to the potential for loss of activity of the glucose analyte prior to the full stated shelf life (August 2018).
Based on the results of internal testing, there is potential for under-recovery of the glucose analyte, and therefore, potential for lower glucose test results. This issue may result in the delay in diagnosis of diabetes. There have been no adverse events reported in conjunction with this product
Recall Action Recall
Recall Action Instructions Point of Care Diagnostics (PTS) is advising users to inspect stock and quarantine any remaing kits of the affected lots. Affected stock is to be destroyed. PTS will issue replacement product or a credit note.

This action has been closed out on the 26/11/2018.
Contact Information 1800 640 075 - Point of Care Diagnostics