Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00251-1
Product Name/Description RPK-01 Luer Connector

Lot Number: MCCC310

Expiry Date:13-10-2021

ARTG Number:145174
(Tekmed - Medcomp Long-term Implantable Haemodialysis Catheter Packs - Catheter, haemodialysis, implantable)
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 28/03/2018
Responsible Entity Tekmed Pty Ltd
Reason/Issue It has been identified that there is a difficulty connecting the female Luer to haemodialysis blood tubing lines, end caps and syringes which may result in blood or air leakage at the connection site.
To date, there have been no reports of injury in Australia as a result of this issue.
Recall Action Recall
Recall Action Instructions TekMed is requesting customers immediately review their stock on hand for affected units and to discontinue their use and distribution. Affected product should be returned to TekMed who will issue a suitable account credit.
New unaffected product is currently being manufactured.

This action has been closed out on 07/11/2018