Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00230-1
Product Name/Description Entroy Pool Lifter

Model Number: GAB1000-x01

Serial Numbers 810000957 up to T3387

ARTG Number: 230595
(Arjo Australia - Lift, patient transfer, stationary)
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 26/03/2018
Responsible Entity Arjo Australia Pty Ltd
Reason/Issue Arjo have identified a potential for the chair or stretcher detachment from Entroy pool lifter. It is identified that this issue is caused by a potential misalignment that might be unnoticed by the caregiver so the seat/stretcher does not get fully paired with the pin. Instead, once the lift reaches certain height, the seat/stretcher falls off the lift. There are one or more possible reasons that can lead to such a situation to occur:

- Maintenance checks not performed properly,
- Users do not recognize the poor connection,
- Pool lift owners do not recognise that the product needs maintenance,
- Wear on the docking connection causes the pin to hit the edge of the pairing cylinder instead of entering it.
Recall Action Product Defect Correction
Recall Action Instructions Arjo are advising users that they will arrange for an on-site service technicians visit to install a Quick Reference Guide, which is an illustration-based instruction, which users may refer to in order to ensure patient transfer is conducted as indicated in the IFU.

This action has been closed out on 15/01/2019
Contact Information 1800 009 077 - Customer Service Centre