Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00159-1 |
Product Name/Description |
Medfusion Syringe Pump Model Series 3500
Product Codes: 3500-0600-50, 3500-0600-51
ARTG Number:122405 (Smiths Medical Australasia Pty Ltd - Infusion pump, syringe) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
23/02/2018 |
Responsible Entity |
|
Reason/Issue |
Smiths Medical has identified that certain Series 3500 Medfusion Syringe Pumps were assembled with a barrel clamp guide containing a ridge within the component. This ridge could potentially lead to spring slippage, resulting in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe being loaded.
|
Recall Action |
Product Defect Correction |
Recall Action Instructions |
A Smiths Medical Representative will be contacting users to arrange repair of affected devices
This action has been closed out on 12/02/2019 |