Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00135-1 |
Product Name/Description |
ProGrade Rel.1x Systems
Product Code: MQB
Model Number: 71209
ARTG Number: 228352 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
12/02/2018 |
Responsible Entity |
|
Reason/Issue |
If the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary. |
Recall Action |
Product Defect Correction |
Recall Action Instructions |
Philips plans to release a new HP firmware to the installed base. In addition the existing Access Point Hardware in the field needs to be patched. A Philips Service Engineer will contact users when the Field Action Kit is available to be implemented.
This action has been closed out on 11/07/2019 |
Contact Information |
1800 251 400 - Philips Customer Care Centre |