Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00125-1
Product Name/Description Foetal Haemoglobin Monoclonal Antibody (HBF-1), FITC. An in vitro diagnostic medical device (IVD)

Catalogue Number: MHFH01

Lot Number: 1626670D

Expiry Date: 30/04/2018

ARTG Number: 212744 - (Life Technologies Australia - Haemoglobin IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/02/2018
Responsible Entity Life Technologies Australia Pty Ltd
Reason/Issue Life Technologies has become aware of a performance issue where the specific lots of Foetal Haemoglobin Monoclonal antibody are displaying high or unusually bright background which could potentially lead to erroneous(false positive) results.
Recall Action Recall
Recall Action Instructions Life Technologies is advising users to discard any remaining product of the affected lot.
A product hold has been placed on future stock, and a credit will be provided for all affected stock.
A review of previously generated test results should be performed at the discretion of the laboratory Director.

This action has been closed out on 07/11/2018
Contact Information 1800 636 327 - Customer Service Support