Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00109-1
Product Name/Description Cortisol Saliva ELISA. An in vitro diagnostic medical device (IVD)

Catalogue Number: RE52611

Lot Numbers: ECO142, ECO143

Expiry Dates: ECO142: 2018-02-28; ECO143: 2018-12-31

ARTG Number: 210636
(Abacus ALS - Clinical chemistry hormone IVDs)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/02/2018
Responsible Entity Abacus ALS Pty Ltd
Reason/Issue The manufacturer, IBL has detected an issue that indicates a degradation in performance of the kit resulting in elevated cortisol values. When testing patient samples with one of the affected lots, the controls and calibrator values will meet specifications; however, due to a degradation in the starting concentration of the calibration standards, the calculated results for individual patient samples may be higher compared to LC-MS/MS measurements and may trigger unnecessary additional testing.

A review of additional data comparing the respective kit standards of the lots to LC-MS/MS data it is evident that the lots ECO142 to ECO143 generate elevated results:

- ECO142 (passing Bablok fit (y=0.009988 + 1.549X)
- ECO143 (passing Bablok fit (y=0.003142 + 1.623X)

There has not been any reports of injuries associated with this issue.
Recall Action Recall
Recall Action Instructions Abacus is advising users to inspect stock and quarantine any remaining stock of the affected batches. Affected product will be replaced with unaffected stock.
A review of previously generated patient results is at the discretion of the Laboratory Director.

This action has been closed out on the 29/10/2018.
Contact Information 1800 222 287 - Customer Service Support