Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00046-1
Product Name/Description CARESCAPE Central Station (CSCS) software version 2.0.2

All-in-one GTIN 00840682109666; Desktop GTIN 00840682109604

ARTG Number: 132673
(GE Medical Systems Australia Pty Ltd - Patient monitor, central unit)
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 12/01/2018
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has become aware of an issue with the CARESCAPE Central Station (CSCS) software version 2.0.2 units. After boot-up or system start up, CSCS V2.0.2 units may display unexpected NO COMM notifications in the Patient Multi-Viewer or Single Viewer window when the CSCS is connected to both the Mission Critical (MC) and the Information eXchange (IX) networks. If a NO COMM notification is displayed, all patient waveforms, parameter data and alarms are lost at the CSCS and patients admitted to Telemetry transmitters are unmonitored unless monitored via other central stations, bedside monitors or direct clinical observation. There have been no injuries reported as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising customers who are experiencing the NO COMM safety issue to follow instructions outlined on the customer letter to restore communication between the CSCS and patient monitoring devices in the interim. GE will arrange for a software correction to resolve this issue.

This action has been closed out on 04/02/2019
Contact Information 1800 659 465 - GE National Call Centre