Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2018-RN-00045-1
Product Name/Description MAQUET VOLISTA central handle on StanOP, TRIPOD and ACCESS Cupolas

Part Numbers: ARD568811950, ARD568811960, ARD568811961, ARD568811962, ARD568812910, ARD568812950, ARD568812960, ARD5688112962, ARD568821913

Serial Numbers (up to, for the above part numbers respectively): 50000, 500063, 500003, 500030, 500004, 50000, 500046, 500007, 500001

ARTG Number: 162037
(Getinge Australia Pty Ltd - Light, surgical)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/01/2018
Responsible Entity Getinge Australia Pty Ltd
Reason/Issue Getinge has received complaints indicating that the central handle holder detaches easily from the cupola during use. We determined that the root cause of this malfunction is excessive tightening of the screws during mounting of the handle holder, resulting in the rivets loosening and detaching from the central handle holder ring.
During manipulation, the handle holder may completely detach from the light head, if all three rivets loosen simultaneously. In this case, although the handle remains in the user’s hand, small quantities of particles may be generated from the plastic holder ring.
Recall Action Product Defect Correction
Recall Action Instructions Getinge will be in contact with affected users to arrange for a service technician to replace existing central holder rings.

This action has been closed out on the 29/11/2018.
Contact Information 1800 438 464 - Getinge Customer Service