Product Name/Description |
MAQUET VOLISTA central handle on StanOP, TRIPOD and ACCESS Cupolas
Part Numbers: ARD568811950, ARD568811960, ARD568811961, ARD568811962, ARD568812910, ARD568812950, ARD568812960, ARD5688112962, ARD568821913
Serial Numbers (up to, for the above part numbers respectively): 50000, 500063, 500003, 500030, 500004, 50000, 500046, 500007, 500001
ARTG Number: 162037 (Getinge Australia Pty Ltd - Light, surgical) |
Reason/Issue |
Getinge has received complaints indicating that the central handle holder detaches easily from the cupola during use. We determined that the root cause of this malfunction is excessive tightening of the screws during mounting of the handle holder, resulting in the rivets loosening and detaching from the central handle holder ring. During manipulation, the handle holder may completely detach from the light head, if all three rivets loosen simultaneously. In this case, although the handle remains in the user’s hand, small quantities of particles may be generated from the plastic holder ring. |