Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00024-1 |
Product Name/Description |
Radiolucent Spine Surgery Frame
Material Number: 100731AO
Model Number: F-70801
Multiple Serial Numbers
ARTG Number: 236335 (Getinge Australia Pty Ltd - Restraint, physical, patient) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
12/01/2018 |
Responsible Entity |
|
Reason/Issue |
Getinge has become aware of a potential risk, which could result in critical harm to the patient if the base of the product were to crack during use. Since the introduction of this product in March 2017, one complaint with no allegation of injury has been reported where the base of the frame cracked as the patient was being prepared for surgery. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Getinge Australia is advising customers to immediately remove the Spine Frame from use that are in the serial number range as identified. Getinge Australia representative will contact users to arrange the corrective repair.
This action has been closed out on the 16/11/2018. |