Type of Product |
Medical Device |
TGA Recall Reference |
RC-2018-RN-00018-1 |
Product Name/Description |
Synchron Enzyme Validator. An in vitro diagnostic medical device (IVD).
Reference Number: 441350
Lot Number: M612600
Expiry: 31/08/2018
ARTG Number: 213974 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
4/01/2018 |
Responsible Entity |
|
Reason/Issue |
Beckman Coulter has identified that a Supplemental Product Information notification is missing from Enzyme Validator kit Lot M612600 due to the introduction of a new lot of traceable reference material in the manufacturing process. As a result, calibration with this Lot may demonstrate differences in AST- or GGT patient and control values compared to calibration with previous lots. Other enzyme assays are not affected. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Beckman Coulter is advising affected customers to refer to the Supplemental Product Information provided with the Customer Letter for adjusting set points of AST- and GGT assays.
This action has been closed out on 25/02/2019 |
Contact Information |
1800 060 881 - Customer Support Centre |