Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01527-1
Product Name/Description Medtronic Clip Gun Scalp Clip System

GUN CG8900 RANEY CLIP KIT 1PK
Model Number: CG8900
Item Number: 00643169109100
Manufacture date: 26/05/2015 - 26/09/2017

KIT CG8091 CLIP GUN 10 PK
Model Number: CG8901
Item Number: 20643169109111
Manufacture date: 26/05/2015 - 02/06/2017

ARTG Number: 98997
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 19/12/2017
Responsible Entity Medtronic Australasia Pty Ltd
Reason/Issue Medtronic Neurosurgery is initiating a recall of all affected model numbers and lot numbers of the Clip Gun Scalp Clip system.

Testing performed by Medtronic showed that the outer sterile pouch of the product did not meet standards for package integrity. Specifically, the testing demonstrated that the outer pouch could be compromised and possibly go undetected by the user which could potentially lead to patient infections.
Recall Action Recall
Recall Action Instructions Medtronic are advising hospitals to cease use of affected product. Medtronic will arrange for retrieval of affected items.

This action has been closed out on 07/11/2018