Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01511-1 |
Product Name/Description |
Prelude SNAP Splittable Sheath Introducer
Vascular Introducer Kit
Multiple Catalogue and Lot Numbers Affected
ARTG 278434 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
22/12/2017 |
Responsible Entity |
|
Reason/Issue |
Merit Medical Systems, Inc. is conducting a recall of specific lots of Prelude SNAP Splittable Sheath Introducers due to an intermittent failure of the splittable hub. Merit Medical has received complaints indicating that the wings on certain lots of the Prelude SNAP have broken off the sheath hub during splitting. This will likely result in a minor delay in procedure. Although rare, this could also result in a pacing lead/catheter displacement, air embolus, or minor hemorrhage. Merit has received no reports of patient harm or injury as a result of this issue. |
Recall Action |
Recall |
Recall Action Instructions |
Users are advised to check inventory for affected devices and discontinue their use. Affected stock should be returned to Merit Medical.
This action has been reviewed and is considered complete in accordance with the requirements of the URPTG. |