TGA Recall Reference RC-2017-RN-01511-1
Product Name/Description Prelude SNAP Splittable Sheath Introducer

Vascular Introducer Kit

Multiple Catalogue and Lot Numbers Affected

ARTG 278434
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/12/2017
Responsible Entity Merit Medical Australia Pty Ltd
Reason/Issue Merit Medical Systems, Inc. is conducting a recall of specific lots of Prelude SNAP Splittable Sheath Introducers due to an intermittent failure of the splittable hub. Merit Medical has received complaints indicating that the wings on certain lots of the Prelude SNAP have broken off the sheath hub during splitting. This will likely result in a minor delay in procedure. Although rare, this could also result in a pacing lead/catheter displacement, air embolus, or minor hemorrhage. Merit has received no reports of patient harm or injury as a result of this issue.
Recall Action Recall
Recall Action Instructions Users are advised to check inventory for affected devices and discontinue their use. Affected stock should be returned to Merit Medical.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.