Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01506-1
Product Name/Description Artis systems with Artis tables

Catalogue number – Artis systems – 5904433, 7008605, 7412807, 7413078, 7555357, 7555365, 7727717, 7728392, 10094135, 10094137, 10094139, 10094141, 10280959, 10848280, 10848281, 10848282, 10848283, 10848353, 10848355, 10848600, 10849000 in combination with various different material number tables.

ARTG: 273952
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/12/2017
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens has been made aware of an issue with the Artis systems with Artis tables. The Artis table top carriage is mounted on top of the table base and fixed with a bolt. A snap right is used to secure the bolt.
In the case of a missing or wrongly installed snap ring the bolt could slide out of its base, and the table top carriage will slide to the floor.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens Healthcare will be in contact with affected customers to arrange for a service to ensure the snap ring has been installed correctly.

This action has been closed out on 18/02/2019
Contact Information 1800 310 300 - Siemens Customer Care