Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01503-1 |
Product Name/Description |
Transseptal Needle
Lot Number: 8181945
Reference Part Number: TSNC-18-71.0
Global Part Number: G02364
ARTG 155227
|
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
13/12/2017 |
Responsible Entity |
|
Reason/Issue |
Cook Medical has become aware that certain Transseptal Needles may be the incorrect length, and is initiating a recall of lot 8181945. Potential adverse events may occur by using the nonconforming products, including being unable to perform the transseptal puncture. This could necessitate replacing the device or rescheduling the procedure. In addition, use of the nonconforming products could make the puncture less controlled, which could potentially lead to injury of the heart. |
Recall Action |
Recall |
Recall Action Instructions |
Cook Medical is advising customers to quarantine affected stock from their inventory for return to the sponsor for a product credit as per the instructions given in the customer letter.
This action has been closed out on 25/01/2021 |