Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01502-1
Product Name/Description Spaplus Weekly wash Protocol and Bottles. An in vitro diagnostic medical device (IVD)

Item Number: TBSIK050S

Lot Number: 415451

ARTG Number: 201721
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 12/12/2017
Responsible Entity In Vitro Technologies Pty Ltd
Reason/Issue The manufacturer, The Binding Site, have identified that the barcode label for the affected lot is incorrect and will not be recognised by the SPAPLUS instrument. This lot number cannot be used to complete the weekly wash. There is no clinical impact associated with this issue
Recall Action Recall
Recall Action Instructions In Vitro Technologies is advising users to quarantine and dispose of any remaining units of the affected product.

This action has been closed out on 6/09/2018.
Contact Information 1300 552 003 - In Vitro Diagnostic