Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01492-1 |
Product Name/Description |
Hercules Patient Lifter
Model Number: H150MT
Serial Numbers: 2015 07 LFT 005T 001 to 2015 07 LFT 005T 079 and 2015 39 LFT 005T 040 to 2015 39 LFT 005T 066
ARTG Number: 224479 |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
8/12/2017 |
Responsible Entity |
|
Reason/Issue |
R&R Healthcare identified that the Hercules Patient Lifter, the spreader bar (which holds the sling) is secured to the lifter arm by a lock tighten dome nut. There has been one report of this dome nut becoming loose causing the spreader to detach from the lifter arm, resulting in a client fall. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
R & R Healthcare is advising distributors to inspect stock and ensure the dome nut is tight. A repair kit will be provided for all lifts to be inspected. Users are advised check that the dome nut is tight, and the spreader bar is secured to the lifter arm. The supplier of the lift will be installing an additional locking pin to the lifter.
This action has been closed out on the 08/11/2018. |
Contact Information |
1800 684 405 - Customer Support |